| FAQ: Possible Problems on Cryosectioning |
 |
 |
 |
| Instrument related problems |
 |
|
|
| 01. |
|
What do I do if the specimen does not advance as expected? |
 |
| |
|
- Check the specimen advance (micrometer) setting to be certain that it is set correctly.
- Activate the coarse specimen advance to confirm that the specimen head is advancing.
- Select trim setting (e.g. 20µm) and try sec read more ...
|
 |
| 02. |
|
How can I remove the specimen disc if it gets stuck or frozen into place? |
 |
| |
|
- Moisture on the underside of the disc may have caused the disc to freeze to the quick-freeze shelf or specimen head. Apply concentrated alcohol (>70%) around the disc, and carefully remove.
- If the disc remains stuck, gentle read more ...
|
 |
| 03. |
|
What do I do if there is no display indication and the cryostat doesn’t work |
 |
| |
|
- Check that the mains lead is connected properly to the outlet and the instrument.
- Check to see if the over-current (GFCI) or circuit breaker has been activated. Reset and switch the cryostat on again with the mains power s read more ...
|
 |
| 04. |
|
What do I do if the chamber (or object) temperature cannot be controlled? |
 |
| |
|
- Ensure that there is a minimum clearance distance of 10 cm on all sides of the instrument.
- Check that the compressor ventilation grids are not dirty or covered. Remove any obstruction or remove dust with brush or vacuum cle read more ...
|
 |
| |
|
|
| |
|
|