FAQ: Leica CV5030 Coverslipper
Air Bubbles
   
01.   Large air bubbles form on the slides – what is the problem?
    A: Reagent incompatibility - e.g. between mounting medium and load container reagent, or between old and new mounting medium.
Check the type of mounting medium and ensure compatibility of all reagents/media used in the system. read more ...
02.   There are small air bubbles on the slides – why?
    A: The nozzle type is not right for the medium dispensed (viscosity of mounting medium either too low or too high for the nozzle type used)
Try another nozzle type.

A: Nozzle is off center
Visually check if med read more ...