FAQ: Leica CV5030 Coverslipper
Coverslips
   
01.   The coverslips are not handled correctly by the instrument – what can be done about this?
    A: The magazine is not inserted correctly
Ensure that the coverslip magazine is filled with coverslips and inserted correctly into the magazine position. Place coverslips flush with one end.

A: Coverslips stick together < read more ...

02.   The coverslips are not properly positioned – what can I do?
    A: The coverslip laydown position needs to be optimized
Go to the corresponding menu - see section 9.3.1 of the Operator's Manual for details. If the coverslips are still not positioned correctly, after the adjustment has been read more ...
03.   Broken coverslips are not detected – what can I do?
    A:There are two possible causes –the coverslip sensor or the suction cups require cleaning or the suction cups are deformed
If the coverslip sensor (located between the red suction cups) appears dirty, clean it with a tissue mo read more ...
04.   Why are broken coverslips thrown into the instrument?
    A: The waste tray needs to be installed.
05.   Why are intact coverslips thrown into the waste tray?
    A:Possibly, the coverslip sensor needs to be cleaned.
Ensure that the sensor is functioning correctly. If the coverslip sensor (located between the red suction cups) appears dirty, clean it with a tissue moistened with xylene (w read more ...
06.   The Pick & Place head collides with the coverslip magazine; a scraping noise can be heard when the coverslip is being picked up – what is wrong?
    A: Most likely, the coverslip magazine is not correctly inserted.
Ensure that the magazine is inserted correctly.
If the problem cannot be remedied, you may have to call a Leica service representative to check the system. read more ...