What do I do if the chamber (or object) temperature cannot be controlled?
|
 |
| FAQ Report |
 |
- Ensure that there is a minimum clearance distance of 10 cm on all sides of the instrument.
- Check that the compressor ventilation grids are not dirty or covered. Remove any obstruction or remove dust with brush or vacuum cleaner.
- If the instrument has been switched off overnight and the automatic defrost cycle did not take place, activate a manual defrost cycle.
- Select a more suitable location for the cryostat if it is exposed to air currents (open windows or doors, overhead air vents).
- Call Technical Service if there might be a leak in the cooling system or if the compressor might be defective.
|
 |
|