What do I do if the chamber (or object) temperature cannot be controlled?
FAQ Report
  1. Ensure that there is a minimum clearance distance of 10 cm on all sides of the instrument.
  2. Check that the compressor ventilation grids are not dirty or covered. Remove any obstruction or remove dust with brush or vacuum cleaner.
  3. If the instrument has been switched off overnight and the automatic defrost cycle did not take place, activate a manual defrost cycle.
  4. Select a more suitable location for the cryostat if it is exposed to air currents (open windows or doors, overhead air vents).
  5. Call Technical Service if there might be a leak in the cooling system or if the compressor might be defective.