| FAQ: Microtomes & Sectioning |
 |
 |
 |
| Instruments related FAQ |
 |
|
|
| 01. |
|
How to avoid a drifting of the hand-wheel? |
 |
| |
|
Spring balance system is not calibrated: Set the hand-wheel to the 3 o'clock position and let it sit. If the wheel drifts upward or downward then the spring system is not balanced and requires adjustment. If the hand-wheel does not drift, t read more ... |
 |
| 02. |
|
Is it possible to adapt the new Leica RM2245 hand-wheel on the RM2145 rotary microtome? |
 |
| |
|
No, it is not possible to fit the RM2245 hand-wheel on the RM2145 rotary microtome. The cover (housing) of the RM2145 has in the area of the hand-wheel a collar which does not fit to the lever of the new hand-wheel brake. Due to the modifie read more ... |
 |
| 03. |
|
The microtome makes "clunking noise" when sectioning |
 |
| |
|
- Linear bearings have moved in cage and are at their lower limits: Linear bearings need to be cleaned, exchanged, possibly re-greased.
- Alignment of bearing cage is incorrect: Open a service call.
- Retraction "ON" read more ...
|
 |
| 04. |
|
No key pad response from control panel |
 |
| |
|
Microtome has no functionality: Check connection on back of microtome for connectivity. Open up a service call if this does not correct the problem. |
 |
| 05. |
|
How can I avoid that the blades wear out very quickly? |
 |
| |
|
The sectioning force is too high: Please adjust the sectioning speed and/or the trimming thickness during sectioning according to specimen hardness. Select a smaller sectioning thickness. |
 |
| 06. |
|
What can I do if there is no further feed motion and thus no sectioning? |
 |
| |
|
- The front end position has been reached: Turn the coarse feed wheel/buttons in the appropriate direction to move the specimen towards the rear limit.
- User may accidentally be holding coarse feed wheel while sectioning. < read more ...
|
 |
| 07. |
|
I am unable to retract the specimen |
 |
| |
|
- The retraction mechanism on the specimen advance system is defective: Open a service call as the spindle may be faulty.
- The PCB control board might be faulty: Open service call to have the control board replaced. Please read more ...
|
 |
| |
|
|
| |
|
|