Leica IP C and IP S cassette/slide printers
     
How can I be sure no flash will be fired when the front print cover is open?
FAQ Report
The IP instruments contain built-in safety devices. In order for the bulb to flash, two serially connected micro-switches must be deperssed by two pins located on the underside of the front cover. If the cover is open, the micro-switches will disengage, the power supply will be disconnected and the bulb will be unable to flash.