| Automating special stains using a commercial Autostainer | |||||||||||||||||||
| Applications 1 | 2 | |||||||||||||||||||
|
|||||||||||||||||||
| (by Rachel Peters, Dept. of Pathology and Laboratory Medicine, Mount Sinai Hospital, 600 University Avenue, Toronto, Ontario M5G 1X5, Canada) The capabilities of the existing hardware and software of the Leica Autostainer XL were explored and some of the hardware was modified: the staining rack adaptor was used to accommodate the staining racks used in this laboratory, and an insert was designed to fit in the autostainer's forced air oven so it could be used for staining purposes rather than as a forced-air slide dryer. |
|||||||||||||||||||
|
|
|||||||||||||||||||